- Office Name: Division of Pharmaceutical Quality Operations IV
- Summary: An FDA inspection of Grifols Diagnostic Solutions Inc. in San Diego, CA, a licensed device/IVD establishment, revealed two significant observations. The firm failed to promptly review, evaluate, and investigate complaints for MDR-reportable events, with technical support staff providing insufficient detail for safety assessments. Additionally, procedures for design changes were found to be inadequately established, specifically concerning the risk assessment for extended onboard stability of licensed test kits and analyzer software.

## Related Documents

- [483 - 2024-06-28](https://www.keypedia.com/records/483/grifols-diagnostic-solutions-inc/8df75157-246b-41ba-9411-17f6920b229a)

## Related Officers

- [Compliance Officer at FDA](https://www.keypedia.com/people/lauren-m-lilly/41582ad0-67b1-4c60-99e2-c1e3c63811c2)
- [Gene D. Arcy](https://www.keypedia.com/people/gene-d-arcy/a2def947-65fb-4056-84e2-1f89cf944cf0)
- [Team Biologics Staff](https://www.keypedia.com/people/linda-thai/f4d9a1c3-e5e8-4f13-8abe-e1f1e597c8f8)

Company: https://www.keypedia.com/companies/grifols-diagnostic-solutions-inc/7fe8042b-3212-4fa9-adaa-05e94b681222

Office: https://www.keypedia.com/offices/division-of-pharmaceutical-quality-operations-iv/a2d2ff91-dc0f-433e-b91f-ca90ff8998b6