# FDA 483 - Karl Storz Endovision, Inc. - September 17, 2021

Source: https://www.keypedia.com/records/483/karl-storz-endovision-inc/fc985e3d-2928-4ded-949a-59af235858f7

> FDA 483 for Karl Storz Endovision, Inc. on September 17, 2021. Product: device. Access full analysis and detailed observations.

---

## Details

- Record Type: 483
- Company Name: Karl Storz Endovision, Inc.
- Inspection Date: 2021-09-17
- Product Type: device
- Office Name: Division of Northeast Imports 
- Summary: An FDA inspection of Karl Storz Endovision, Inc. in Charlton, MA, a medical device manufacturer, revealed significant deficiencies in design control processes. The firm failed to establish a design history file for its 11101VNS Strobo-Video-Rhino-Laryngoscope, lacked adequate design change procedures, and did not properly verify design outputs against input requirements. These issues are serious as they are linked to complaints of operator burns and high heat associated with the device.

## Related Documents

- [WARNING_LETTER - 2014-10-29](https://www.keypedia.com/records/warning_letter/karl-storz-endovision-inc/eacae5de-3786-48e7-975d-7a076004b4bc)
- [EIR - 2021-09-17](https://www.keypedia.com/records/eir/karl-storz-endovision-inc/5eb239d0-52dd-4647-8747-277e4b9a0675)
- [483 - 2022-06-14](https://www.keypedia.com/records/483/karl-storz-endovision-inc/f7c2b3fb-1f4f-49c6-8575-a100157c415c)

## Related Officers

- [Power Trader/Data Scientist](https://www.keypedia.com/people/jeffrey-j-thibodeau/62b74030-4aef-400b-8ae2-8bf84712d813)

Company: https://www.keypedia.com/companies/karl-storz-endovision-inc/ccba1a5b-ab36-46d6-aa92-dbd2fab934b7

Office: https://www.keypedia.com/offices/division-of-northeast-imports/b20599cf-9797-41fd-947f-6589ef596a94