FDA 483 - Abbott Diagnostics Scarborough, Inc. - March 02, 2018
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Alere Scarborough, Inc. in Scarborough, ME, was cited for significant deficiencies in its Medical Device Reporting (MDR) processes during an FDA inspection. The firm failed to submit timely MDRs for false positive and false negative results from its Alere Determine HIV-1/2 Ag/Ab Combo assay, which could cause or contribute to serious injury or death. Additionally, the company's written MDR procedure was found to be inadequate for identifying and evaluating reportable events.
ID · afcdce87-41b8-4e97-959e-259375e2fc16