FDA 483 - Abbott Medical - June 28, 2023
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Abbott Medical in Sylmar, CA, a medical device manufacturer, revealed significant deficiencies in their quality system. The firm failed to report corrections and removals for implantable cardiac devices with Bluetooth communication issues that could lead to serious injury or death. Additionally, procedures for corrective and preventive actions and complaint handling were found to be inadequate, with numerous complaints misclassified and not properly evaluated for Medical Device Report (MDR) reportability.