FDA 483 - Amrex Zetron Inc - February 08, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Amrex Zetron Inc, a medical device manufacturer in Paramount, CA, was cited for multiple significant quality system deficiencies during an FDA inspection. Key issues include failures related to Unique Device Identification (UDI) for their Class II muscle stimulators and ultrasound devices, inadequate device history records, and a lack of established procedures for quality audits. Additionally, the firm failed to properly process service reports as complaints, even when they described serious device malfunctions like electrocution, with some issues being repeat observations.
ID · 7e03d2e4-63fe-4675-bf29-6f727d4e1600