483
B Positive National Blood ServicesFDA 483 - B Positive National Blood Services
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
An FDA inspection of B Positive National Blood Services in Easton, Pennsylvania, from November 5-7, 2024, revealed two significant observations. The firm failed to recalibrate a critical device used for quality control checks of collection equipment and did not conduct a required performance qualification study for freezers used to store Source Plasma units prior to their use.