FDA 483 - Bausch & Lomb Incorporated - May 15, 2006
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA Form 483 details numerous observations from an inspection related to the manufacturing and quality systems for ReNu w/ MoistureLoc Multi-Purpose Solution.
**Facility and Operations:** The inspection covers manufacturing processes for contact lens solutions, including ReNu w/ MoistureLoc, Sensitive Eyes, and Boston Cleaner.
**Violations and Observations:**
* **Design Control Deficiencies:** The design plan for ReNu w/ MoistureLoc was incomplete, lacking descriptions of activities, responsibilities, and interfaces. A complete design plan was not done, and traceability was unclear. Formal feasibility and risk assessment were not performed after combining two preservative formulations in 2004. * **Specification and Testing Issues:** Raw material specifications were not firmly established prior to process validation. Analytical in-process and finished product specifications were not assigned in the design plan or firmly established before launch; for example, the preservative specification was lowered post-process validation. No test method exists to evaluate Alexidine degradation. * **Design History File and Reviews:** The design history file lacked an R&D readiness statement. Incomplete design input requirements were not addressed, with several outstanding for ReNu w/ MoistureLoc prior to market launch. Post-launch product reviews for ReNu w/ MoistureLoc were not performed as required by established procedures (BLR-POL-400), despite the product being distributed since August 2004. * **MDR and Recall Reporting Fail
ID · 7d50142b-ed55-41e3-b196-18edb1340d4e
Full citation text and observation details available on the Dashboard.