Baxter Healthcare CorporationBaxter Healthcare Corporation – FDA 483, Aug 18, 2023
Record Details
An FDA inspection of Baxter Healthcare Corporation in Round Lake, IL, identified that procedures for design validation were not adequately established. Specifically, a claim regarding the Novum IQ Syringe Pump's Smart Battery Pack degradation was not validated using production-equivalent units. This indicates a deficiency in ensuring product claims are thoroughly tested according to established procedures.