FDA 483 - Busse Hospital Disposables, Inc. - April 13, 2018
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Buerer Hospital Disposables, Inc. in Hauppauge, NY, a manufacturer of hospital disposables, revealed two significant issues. The firm failed to adequately establish procedures for monitoring and controlling process parameters for a validated process, specifically noting identical Installation Qualification and Operation Qualification documents for a towel machine. Additionally, the firm did not adequately establish requirements for suppliers and contractors, as evidenced by a failure to open a CAPA for multiple vendor non-conformances related to syringes.
ID · c87d1fc2-445d-48a0-88e6-3acf3c017412