CareFusion 303, Inc. – FDA 483, Apr 6, 2020
CareFusion 303, Inc. in San Diego, CA, a medical device manufacturer, was inspected by the FDA from March 2 to April 6, 2020. The inspection revealed significant deficiencies across multiple quality system areas, including inadequate corrective and preventive actions, complaint handling, medical device reporting, design verification, servicing procedures, and software validation. These issues were linked to numerous device malfunctions, some resulting in patient deaths or serious injuries, and included a failure to report a major field correction to the FDA.
ID · 6fb77e5f-514f-4201-a4d9-a5cc5866ed99
Full citation text and observation details available on the Dashboard.
Comes with the profile
KRA (KeyPedia Risk Analysis)
How likely the FDA is to take enforcement action in the next 12 months.
Example18-month trend