FDA 483 - Catalent Belgium SA - August 19, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
This FDA Form 483 details multiple deficiencies observed during an inspection of a facility manufacturing drug products for the US commercial market, including a combination drug/device.
Key observations include: - **Aseptic Processing Areas:** Air visualization studies in RABs failed to demonstrate sufficient airflow velocity over critical areas of filling machines, particularly during interventions. - **Deviation Investigations:** The firm failed to thoroughly review unexplained discrepancies and perform proper root cause analysis for sterilizing equipment issues, despite recurring observations of sterilized part contamination between 2019 and 2021. - **Process Control:** Aseptic filling process simulations (media fills) performed after significant HVAC changes were deficient, failing to consider major changes to HEPA filters and not simulating worst-case conditions (larger openings, low filling speed). - **Equipment Qualification:** - Sterilization cycle requalification used incorrect part types. - Integrity tester performance qualification studies were deficient, with incorrect study parameters and potential for false negatives. - Integrity test results were recorded as Pass/Fail without numerical values or initial/final pressure records. - Equipment clean hold time studies were deficient, not demonstrating cleanliness after a single cleaning cycle. - Climate-controlled chamber qualifications (stability chambers, refrigerators, incubators) were deficient, lacking recorded temperature/humidity recovery times. - **Data Integrity:** - QC instruments and equipment lacked protection from manipulation; procedures allowed analysts to change date and
- Company
- Catalent Belgium SA
- Inspection Date
- August 19, 2022
ID · a5641d00-652f-47a6-a7f2-a1b291a6e2c8
Full citation text and observation details available on the Dashboard.