CooperSurgical, Inc. – FDA 483, Jul 15, 2021
CooperSurgical, Inc. in Trumbull, CT, a medical device manufacturer, was cited for significant deficiencies during an FDA inspection. Observations included uncontrolled labeling and packaging operations leading to multiple product recalls, failure to submit Medical Device Reports within the required 30-day timeframe, and inadequate procedures for handling nonconforming products. Additionally, the firm lacked established procedures for controlling environmental conditions, specifically clean room wall cleaning.
ID · 1674aa4d-27cb-46d5-81e2-559b388a3366
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.