FDA 483 - CooperSurgical, Inc. - February 01, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
CooperSurgical, Inc. in North Tonawanda, NY, a manufacturer of drug/medical device combination products, was cited with two observations during an FDA inspection. The inspection revealed significant deficiencies in the firm's corrective and preventive action (CAPA) procedures, including inadequate investigations and lack of verification for corrective actions. Additionally, the firm failed to adequately validate its pouch sealing process, lacking proper statistical rationale for sample sizes.
ID · b95fba02-cef5-4e8c-a8d7-e19e31e5a0dd