483
•CooperVisision Manufacturing Puerto Rico LLC•April 25, 2008

CooperVisision Manufacturing Puerto Rico LLC – FDA 483, Apr 25, 2008

Record Details

This FDA Form 483 documents observations made during an inspection of a firm on April 25, 2008. The inspection revealed three key observations related to quality system requirements.

Observation 1 notes that process validation activities and results for the sterilization of contact lenses in single blisters were not fully documented. Specifically, deviations from pre-established acceptance criteria during the performance qualification protocol for the (b)(4) Autoclave, conducted in July 2002, were not fully reviewed or documented in the final report. Discrepancies found in raw data, but not in the final report, include replacement of temperature indicators, execution of an additional sterilization cycle, an additional cycle for biological indicator testing, and a temperature failure (120 Deg) in one chamber's temperature indicator during "Run 1 Max," with no biological indicator results for this cycle. The firm promised to correct this.

Observation 2 states that complete complaint files are not maintained. The firm failed to fully document the evaluation of the manufacturing process for locally manufactured product lots during investigations of consumer complaints reporting patient reactions from 2005 to July 2007. These evaluations did not comprehensively assess manufacturing records to determine the impact of reported inspection results or other conditions on the reported defects. The firm reported this as corrected, but it was not verified.

Observation 3 indicates that procedures for implementing corrective and preventive actions (CAPA) were incomplete. Trend Procedure #0

Company
CooperVisision Manufacturing Puerto Rico LLC
Inspection date
April 25, 2008
Product type
Devices
Office
San Juan District Office
Person
  • Noreen Muniz (Investigator)
View Company Profile

ID · cc04f6dc-70e5-45eb-bce5-42bfab043366

Violation Codes3
21 CFR 820.19821 CFR 820.10021 CFR 820.168

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore CooperVisision Manufacturing Puerto Rico LLC
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.