DePuy Orthopaedics, Inc. – FDA 483, Sep 21, 2004
On September 21, 2004, FDA Investigator Robert G. R:tiff issued a Form FDA-483 to Depuy Orthopaedics, Inc., a medical device manufacturer located at PO Box 988, 700 Orthopaedic Drive, Warsaw, IN 46581-0988. The inspection, conducted from September 13, 2004, to September 21, 2004, identified two observations related to the firm's quality system.
Observation 1 noted that procedures to control the design process of the device were incomplete. Specifically, risk analysis procedures did not document the mechanism by which unacceptable risks are determined. The firm reported this issue as corrected, but it was not verified by the investigator.
Observation 2 stated that procedures for defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements were not complete. More specifically, procedures used to identify design outputs essential to the proper functioning of devices designed by the firm had not been documented. This observation was also reported as corrected by the firm but not verified.
- Office
- Detroit District Office
- Person
ID · 5ccaf419-ddfc-4332-9770-75b1f9a4e265
Comes with the profile
KRA (KeyPedia Risk Analysis)
How likely the FDA is to take enforcement action in the next 12 months.
Example18-month trend
Form 483 · September 21, 2004
What you get
A document
One file.
- One FDA document, such as a Form 483
- Yours for life in your Library
- Each has its own price
The profile
All the documents, plus intelligence about the company.
- Every DePuy Orthopaedics, Inc. document on file
- The KRA score, which a single document does not include
- Observations and inspection history
Access
This, for every company on KeyPedia.
- Combine profiles in one search
- Search across any modality: products, industry types, drugs, devices and recalls
- Track everything as it changes
- Other agencies too, such as EMA, Health Canada and UK MHRA