Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. 483
  6. /
  7. Dispo-Med (Malaysia) Sdn Bhd
483
•Dispo-Med (Malaysia) Sdn Bhd•June 15, 2023

Dispo-Med (Malaysia) Sdn Bhd – FDA 483, Jun 15, 2023

Record Details

An FDA inspection of Dispo-Med (Malaysia) Sdn Bhd, a medical device manufacturer, was conducted from June 12-15, 2023, resulting in a Form FDA 483 listing four key observations related to deficiencies in their quality management system. The company's procedures for Corrective and Preventive Actions (CAPAs) were found inadequate, specifically lacking documented investigations for root causes and corrective actions for recurring nasal prong component rejections.

Design Change procedures were also deficient. A filter modification for a patient monitoring system was implemented without technical justification or verification testing. Furthermore, a material change for nasal prongs lacked proper documentation, approval, and a required risk assessment. Procedures for accepting incoming products were not adequately established, as Certificates of Conformity from suppliers failed to confirm essential blockage testing for Airway Adapters, despite prior commitments.

Lastly, production processes for the nasal prong CO2 cannula product were not properly controlled. The dipping process lacked adequate control over temperature and timing, and relevant devices were uncalibrated. An on-site observation revealed dipping times outside specified ranges, and the Operational Qualification did not establish an acceptable process parameter range.

These observations underscore the need for Dispo-Med to enhance its quality system to ensure product specifications are consistently met. The firm has committed to addressing these issues within six months.

Company
Dispo-Med (Malaysia) Sdn Bhd
Inspection date
June 15, 2023
Product type
devices
Office
Office of Regulatory Affairs
Person
  • Stephen D. Eich (investigator)
View Company Profile

ID · 11e70f17-611e-44cc-ae60-b0e598eb0e01

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Dispo-Med (Malaysia) Sdn Bhd
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.