FDA 483 - Dr Reddy's Laboratories Limited - December 08, 2023
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
This FDA Form 483 document details deficiencies observed during an inspection related to laboratory records, data integrity, and employee training.
**Facility and Operation Details:** The inspection pertains to a facility producing (b)(4) Injection and (b)(4) API. The firm conducted a sameness study for (b)(4) Injection to compare it to a Reference Listed Drug (RLD) using Circular Dichroism Spectropolarimeter (equipment ID: AA266) and Nuclear Magnetic Resonance Spectrometer (NMR, equipment ID: AA574).
**Violations and Observations:** * **Deficient Laboratory Records:** Laboratory records for the sameness study of (b)(4) Injection were found to be deficient, lacking complete and accurate data. * **Omitted and Unreported Electronic Data:** * An initial Circular Dichroism spectra scan for Test Drug batch (b)(4) on 11/25/2016, created at (b)(4), was omitted from analysis. A second scan was reported instead, and a temperature fluctuation incident was not investigated or reported to the agency. * Circular Dichroism data for RLD batch #(b)(4) and Test Drug batch (b)(4) generated on 11/28-29/2016 was not reported and invalidated without scientific justification. A Laboratory Investigation Report (LIR-A-AD