483
DRW MedicalFDA 483 - DRW Medical - April 07, 2025
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
DRW Medical, a manufacturer in Aston, PA, was inspected regarding its Class II NAPA LP-15 Airway Pressure monitor. The inspection revealed significant deficiencies in the firm's design control process, including a failure to complete required design input, output, and transfer checklists. Additionally, the firm lacked adequate justification or validation for software changes made to the device, raising concerns about its safety and effectiveness for neonates.
- Person
Open in Dashboard
ID · f14f0cfd-1323-4abf-9061-370c3b1a9c99