FDA 483 - Inspire Medical Systems Inc. - May 10, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Inspire Medical Systems Inc. in Golden Valley, MN, was cited for four observations during an FDA inspection. The firm failed to adequately establish procedures for purchasing controls, supplier management, and corrective and preventive actions. Additionally, complaint investigations for potential device failures, especially infections, were not adequately conducted, and procedures for monitoring process parameters for validated processes were inadequately defined and implemented.
ID · 91307810-95e3-4c11-8809-dc5bfa6ab003