483
JAS Diagnostics, Inc./Drew Scientific, Inc.FDA 483 - JAS Diagnostics, Inc./Drew Scientific, Inc. - August 29, 2019
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
JAS Diagnostics/Drew Scientific, Inc./Diamedix Corp. in Miami Lakes, FL, was inspected and cited for significant deficiencies across its quality system for medical devices. The inspection revealed inadequate organizational structure, incomplete corrective and preventive actions, missing design history files, and insufficient risk analysis and process validation. Several observations were repeats from previous inspections, indicating a persistent lack of compliance.
Open in Dashboard
ID · 7159ca57-fcb9-4a5d-bf03-475f5a86ff6d