Linde Gas & Equipment Inc – FDA 483, Jul 25, 2025

Record Details

**FDA 483 Summary: Linde Gas & Equipment Inc**

**Inspection Dates:** July 15-25, 2025 **Location:** Livingston, TN **FEI Number:** 3012160128 **Report Issued to:** John E. Holland, Director of Healthcare Operations

**Key Violations:**

1. **Corrective and Preventive Action (CAPA) Deficiencies:** Linde Gas & Equipment Inc failed to adequately establish procedures for corrective and preventive actions, particularly concerning the NOxBOXi, a Class II medical device. The device is crucial for delivering and monitoring nitric oxide therapy in neonates.

2. **Recurrent Device Malfunctions:** The firm has not effectively addressed recurring malfunctions, such as NO2 sensor failures, which can lead to inadequate nitric oxide delivery and undetected NO2 toxicity. These issues are linked to software discrepancies.

3. **Off-label Use:** Despite identifying off-label use of the NOxBOXi, which resulted in adverse events including one death, the firm’s corrective actions, such as reviewing clinician training materials, were deemed insufficient.

4. **Dose Fluctuations:** The NOxBOXi exhibited dose fluctuations during High Frequency Jet Ventilation, compromising treatment efficacy. The firm communicated this issue to the manufacturer but did not implement further corrective actions.

5. **Internal Fault Alerts:** Devices issued alerts during therapy, disrupting treatment and causing oxygen desaturation in neonates. Although a software solution was identified, implementation has been delayed due to resource constraints.

**Regulatory Framework:** The firm is responsible for compliance with FDA quality system requirements, which mandate internal audits and timely corrective actions to address any violations.

**Required Actions:** Linde Gas & Equipment Inc must develop and implement effective corrective actions to address the identified deficiencies, particularly focusing on software improvements and ensuring compliance with device specifications to mitigate risks to patient safety.

Company
Linde Gas & Equipment Inc
Inspection date
July 25, 2025
Product type
Device
Office
New Orleans District Office
Person
  • Jamie P. Webb
View Company Profile

ID · 79091805-cfd9-4695-b0b0-93fe7fd0328f

Violation Codes4
21 CFR 806.1021 CFR 820.181(s)21 CFR 820.5021 CFR 806.10(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Linde Gas & Equipment Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.