483
Medical Solutions International, Inc.FDA 483 - Medical Solutions International, Inc. - April 15, 2024
Discuss this record with AI
Record Details
Medical Solutions International, Inc. in Merriam, KS, a manufacturer of Class II medical devices, was cited for significant deficiencies in its design control procedures. The inspection revealed inadequate establishment of procedures for design validation, incomplete risk analysis, and insufficient procedures for both design verification and design change. All four observations were noted as repeat findings from a previous FDA inspection.
Open in Dashboard