483
Medtronic MiniMed, Inc.FDA 483 - Medtronic MiniMed, Inc. - July 22, 2016
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
Medtronic MiniMed Inc. in Northridge, CA, was inspected from July 5-22, 2016, revealing issues with device accountability. Specifically, records for the disposition of GST4C transmitters were not accurate or complete, leading to the use of incorrect software versions in a clinical study. This discrepancy was not identified until after clinical data was submitted to the FDA for IDE and PMA review.