•Medtronic Puerto Rico Operations Co.•June 13, 2006

Medtronic Puerto Rico Operations Co. – FDA 483, Jun 13, 2006

Record Details

The FDA conducted an inspection of Medtronic Puerto Rico Operations Co., a medical device manufacturer located in Juncos, Puerto Rico, from March 13 to June 13, 2006. The inspection revealed several significant compliance issues related to production and process controls, which were documented in the FDA Form 483.

Key violations included inadequate validation of software used in the manufacturing of Paradigm external insulin pumps. The software validation failed to demonstrate that it met specified requirements, lacked testing for certain error messages, and did not provide written justifications for omitting specific tests. Additionally, the validation process for the automated adhesive dispensing and assembly processes was found lacking, as it did not adequately address equipment setup failures or the impact of reduced sample sizes on validation outcomes.

Further observations noted that certain test equipment was unsuitable for its intended purposes, as it did not ensure consistent pressure application during sensor calibration. Moreover, procedural deviations were identified, such as the failure to generate required deviation authorizations for non-conforming torque tubes.

The FDA also highlighted deficiencies in the verification and validation of corrective and preventive actions (CAPA). Specifically, there was no documented evidence to confirm that new preventive maintenance procedures would effectively prevent calibration failures.

Medtronic is required to address these issues by conducting thorough internal audits to identify and rectify all violations of quality system requirements, ensuring compliance with FDA regulations.

Company
Medtronic Puerto Rico Operations Co.
Inspection date
June 13, 2006
Product type
Devices
Office
San Juan District Office
Person
  • Lisa M. Lopez
View Company Profile

ID · b9e674e9-2c3e-4fbe-82c6-2d10a3b186eb

Other documents for Medtronic Puerto Rico Operations Co.
  • 483 - 2019-10-25October 25, 2019
  • EIR - 2019-10-25October 25, 2019
  • 483 - 2019-02-21February 21, 2019
  • EIR - 2018-05-15May 15, 2018
  • 483 - 2018-05-15May 15, 2018

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Medtronic Puerto Rico Operations Co.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.