483
NIPRO CORP.FDA 483 - NIPRO CORP. - July 05, 2018
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
An FDA inspection of Nipro Corporation - Odate Factory in Odate, Japan, revealed significant deficiencies in process validation for their PATHFAST® cTnI reagent cartridge kits. The firm failed to adequately validate manufacturing processes, citing issues with single-lot validation, insufficient sample size justification, and the absence of an Operational Qualification. These findings indicate a lack of robust quality control in their medical device manufacturing.
Open in Dashboard
ID · 92df3935-1f99-4f04-af85-e76749faed66