483
Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.FDA 483 - Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc. - February 15, 2019
Discuss this record with AI
Record Details
An FDA inspection of Polymer Technology Systems, Inc. in Indianapolis, IN, a medical device manufacturer, revealed significant quality system deficiencies. The firm failed to adequately validate processes and designs, maintain effective corrective and preventive actions, and ensure proper supplier controls. Additionally, the company did not submit required Medical Device Reports for serious malfunctions, including battery explosions, and lacked proper health impact assessments for product corrections.
- Inspection Date
- February 15, 2019
- Product Type
- Device
Open in Dashboard
ID · 6fb81136-f4a9-4b04-98f4-5eda07909f52