FDA 483 - Scientific Analytics/DARI Motion, Overland Park, KS USA - August 26, 2021
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of DARI Motion, Inc. in Overland Park, KS, a manufacturer of the Class II DARI Health Interactive Rehabilitation Exercise system, revealed significant deficiencies across its quality management system. The firm failed to establish or adequately maintain fundamental procedures for design control, corrective and preventive actions, device master records, acceptance activities, complaint handling, quality audits, training, and management review. These widespread issues indicate a severe lack of compliance with medical device quality system regulations.
- Inspection Date
- August 26, 2021
- Product Type
- Device
ID · ab3814f3-77d2-4c89-b7a5-210c3a741064