FDA 483 - Shape Memory Medical, Incorporation - March 13, 2025
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Shape Memory Medical, Incorporation in San Jose, CA, received a Form 483 for inadequate procedures related to receiving, reviewing, and evaluating complaints for its Class 2 IMPEDE FX device. The inspection found that the firm's complaint handling procedure failed to provide sufficient guidance on managing investigations, documenting investigator actions, and ensuring timely processing and escalation of complaints. Multiple complaints were closed outside of established timelines without proper documentation or escalation.
ID · 3989f59b-5a60-450d-b002-4706e7b4dd40