FDA 483 - Siesta Medical, Inc. - December 15, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Siesta Medical, Inc. in Los Gatos, CA, was cited for significant deficiencies in its quality system related to its Encore System medical device. The inspection revealed failures in timely Medical Device Reporting, inadequate complaint handling procedures and investigations, insufficient design change controls, and incomplete risk analysis for identified device failure modes. These issues indicate a systemic breakdown in ensuring device safety and regulatory compliance.
ID · 7ccf55a6-dc9f-415c-a1d8-542032217644