Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. 483
  6. /
  7. Stryker Corporation
483
•Stryker Corporation•May 14, 2009

Stryker Corporation – FDA 483, May 14, 2009

Record Details

The FDA Form 483 inspection of Stryker Instruments revealed multiple violations related to quality systems and manufacturing processes for various medical devices, including Clover Catheters, Silver Ex-Fen Catheters, Stryker Auto-Fuser PainPumps, and other surgical instruments.

**Key Violations and Observations:**

* **Inadequate Process Validation:** * **Clover and Silver Ex-Fen Catheters:** The injection molding process for the female luer hub was not fully validated. Validation runs showed dimensional failures, short shots, flash, and contamination, but were not repeated with revised parameters. Nominal runs were not representative of routine production duration. The extrusion process for tubing was also not validated, with only outer diameter continuously monitored. * **Stryker Auto-Fuser PainPump:** The bonding process of the glass restrictor to the male luer was not completely validated. Limited test runs were performed, but no documentation of nominal runs or destructive cross-sectional examination of the bond quality was available. This bonding issue was identified as the root cause of an over-infusion malfunction.

* **Failure to Submit MDRs:** Multiple complaints received between January 23, 2009, and February 18, 2009, reporting leakage of fluid or black oil/substance from surgical instruments into the sterile field or patient wound were not reported as Medical Device Reports (MDRs) within 3

Company
Stryker Corporation
Inspection date
May 14, 2009
Product type
Devices
Office
Detroit District Office
Person
  • Charles Spyr (Real Estate Investor)
View Company Profile

ID · 32ef39f2-6cb4-4898-9b77-a198e617dd19

Violation Codes10
FD&C Act 801(a)(3)FD&C Act 510(k)FD&C Act 502FD&C Act 50121 CFR 820.9021 CFR 80121 CFR 820.72FD&C Act 80321 CFR 820.25FD&C Act 820

Full citation text and observation details available on the Dashboard.

Other documents for Stryker Corporation
  • WARNING_LETTER - 2012-11-09November 9, 2012

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Stryker Corporation
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.