Attention www.globalkeysolutions.net is now www.keypedia.com, read the full press release here.
KeyPedia LogoKeyPedia Icon
  1. Home
  2. /
  3. Records
  4. /
  5. 483
  6. /
  7. TENTAMUS INDIA PRIVATE LIMITED
483
•TENTAMUS INDIA PRIVATE LIMITED•August 22, 2025

TENTAMUS INDIA PRIVATE LIMITED – FDA 483, Aug 22, 2025

Record Details

An FDA inspection of Tentamus India Private Limited, a contract testing laboratory in Hyderabad, Telangana, India, conducted from August 14 to August 22, 2025, revealed significant data integrity and quality control deficiencies. The FDA issued a Form 483, citing multiple observations indicating potential violations of Good Manufacturing Practices.

Key issues included a systemic failure within the Quality Control unit, evidenced by the destruction of critical GMP documents, including chromatograms and raw data, and employees attempting to hide and backdate incomplete test documents. Investigators found torn GMP documents in trash bags and observed employees using personal diaries to record initial testing information, leading to adjustments of metadata to meet passing limits. There was also a lack of adequate oversight on the issuance and reconciliation of GMP forms and analytical worksheets.

Furthermore, the firm failed to adequately investigate unexplained discrepancies, often correcting issues informally without root cause analysis or corrective actions. Examples included reissuing test reports with "Too Numerous To Count" results without investigation and using an expired reference standard for over a year and a half in numerous analyses. A critical digital colony counter, essential for microbial plate colony counts, had non-operational lights for approximately seven years without repair, leading to potential inaccuracies.

Repeat observations from a September 2022 inspection highlight a persistent failure to follow established laboratory control mechanisms, including delayed and non-contemporaneous documentation, and backdating records. The company's leadership also provided misleading information regarding 'Routine Process Audit' reports. These findings necessitate comprehensive corrective and preventive actions to ensure data integrity, adherence to procedures, and reliable testing results.

Company
TENTAMUS INDIA PRIVATE LIMITED
Inspection date
August 22, 2025
Product type
drugs
Office
Office of Inspections and Investigations
Person
  • Pratik S. Upadhyay
View Company Profile

ID · a64623e5-2b62-4f6c-a980-85528906fc5d

Other documents for TENTAMUS INDIA PRIVATE LIMITED
  • WARNING_LETTER - 2025-08-22August 22, 2025
  • 483 - 2025-08-14August 14, 2025
  • 483 - 2022-09-23September 23, 2022
  • 483 - 2019-11-01November 1, 2019
  • 483 - 2016-07-15July 15, 2016

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore TENTAMUS INDIA PRIVATE LIMITED
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.