•Terumo Cardiovascular Systems Corporation•March 29, 2010

Terumo Cardiovascular Systems Corporation – FDA 483, Mar 29, 2010

Record Details

This FDA Form 483 documents observations from an inspection, highlighting significant deficiencies in the facility's quality system and manufacturing processes. Management with executive responsibility failed to ensure the full implementation and maintenance of an adequate and effective quality system across all organizational levels.

Key violations include: - Inadequate investigation of nonconformities to determine root causes. - Failure to identify all necessary corrective and preventive actions (CAPA) to prevent recurrence of nonconforming products and quality issues. - Lack of verification or validation of CAPA to ensure effectiveness and prevent adverse effects on finished devices. - Absence of defined and implemented CAPA procedures utilizing appropriate statistical methodology for identifying existing and potential causes of nonconformities. - Insufficient documentation of CAPA activities, including verification/validation. - Unimplemented complaint handling procedures for receiving complaints. - Lack of established quality system procedures.

Regarding medical device reporting (MDR), the facility failed to: - Submit an MDR report within 30 days of becoming aware of information suggesting a marketed device caused or contributed to a death or serious injury. - Provide a complete description of the event, device involvement, and problem nature in Block B of FDA Form 3500A. - Submit supplemental reports within one month of receiving new information not included in initial reports.

Further issues include: - Failure to establish adequate quality requirements for suppliers. - Inadequate validation and approval of processes whose results cannot be fully verified

Company
Terumo Cardiovascular Systems Corporation
Inspection date
March 29, 2010
Office
Detroit District Office
People
  • Ryan Zettle, PharmD, MBA
  • Charles Spyr (Real Estate Investor)
  • Kimberly Lee
  • Gary D. Urbiel Goldner (Investigator )
View Company Profile

ID · 91358452-97b4-4c1b-8281-77aced3673c6

Violation Codes8
FD&C Act 801(a)(3)21 CFR 820.5021 CFR 820.19821 CFR 820.10021 CFR 820.3021 CFR 820.7521 CFR 820.2021 CFR 803.33

Full citation text and observation details available on the Dashboard.

Other documents for Terumo Cardiovascular Systems Corporation
  • 483 - 2012-09-28September 28, 2012
  • 483 - 2008-08-11August 11, 2008
  • 483 - 2005-12-09December 9, 2005
  • 483 - 2004-05-18May 18, 2004

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Terumo Cardiovascular Systems Corporation
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.