483
Unetixs Vascular, Inc.FDA 483 - Unetixs Vascular, Inc. - June 11, 2018
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
OneLuxe Vascular Inc., a manufacturer in North Kingstown, RI, was cited for significant quality system deficiencies during an FDA inspection. The firm failed to establish adequate procedures for corrective and preventive actions, properly investigate device complaints, and ensure timely review of service records. All three observations were repeats, indicating persistent non-compliance with regulatory requirements.
Open in Dashboard
ID · 97d4301b-79fe-4c2d-8da4-30eb07529925