China NMPA Product Recall - Soft (hydrophilic) contact lens
Discuss this record with AI
On January 13, 2023, CooperVision Inc., through its entity CooperVision Products Trading (Shanghai) Co., Ltd., initiated a voluntary recall of specific batches of its Soft (hydrophilic) Contact Lenses. This action was prompted by the discovery of discrepancies between the power prescription printed on the product labels and the actual power of the lenses contained within certain models and production batches. The recall, classified as Class III by the National Medical Products Administration (NMPA), addresses a critical issue that could potentially impact user experience and vision correction if patients receive lenses with incorrect specifications. While no specific inspection dates were provided, the recall announcement itself signifies proactive compliance with regulatory standards. CooperVision Inc. is undertaking this voluntary recall to ensure product quality and patient safety, instructing the collection of affected products. Further details regarding the specific models, specifications, and batch numbers subject to this recall are documented in the accompanying "Recall Event Report Form." This proactive measure underscores the company's commitment to rectifying product anomalies and upholding consumer trust in medical device integrity.
ID · 7ca5033c-1542-471b-9942-2198f26f9687