China NMPA Product Recall - VentStar disposable pediatric ventilation circuit
Discuss this record with AI
Dräger Medical Equipment (Shanghai) Co., Ltd. has initiated a voluntary Level I recall for specific disposable pediatric ventilation circuits used with Oxylog 3000 and Oxylog 3000 Plus ventilators. This action follows internal testing that revealed a leak in the one-way valve of the VentStar Oxylog3000F(P)190 disposable pediatric ventilation circuit, part number 5704964, affecting products manufactured up to March 2016. The identified defect could allow exhaled air to re-enter the breathing circuit, potentially causing patients to repeatedly inhale exhaled air and experience a dangerous reduction in oxygen concentration. Under the regulatory framework of the National Medical Products Administration (NMPA), Dräger has implemented several required actions. These include immediately halting the distribution of all affected stock and managing its disposal. The manufacturing facility has also enhanced its quality control by implementing 100% leak testing for all newly produced circuits. Additionally, the company has issued safety notices to customers, instructing them on the proper handling and return of the recalled materials. Although no complaints or adverse events have been reported to date, this proactive recall addresses a significant potential safety risk to patients.
- Company
- Draegerwerk AG & Co. KGaA
ID · 890ace93-b247-4650-a22b-1c97becf0f46