China NMPA Product Recall - Hip joint femoral neck component
Discuss this record with AI
The National Medical Products Administration (NMPA) issued a recall notice on August 12, 2015, concerning a voluntary recall initiated by Microport Orthopedics, Inc., with its responsible unit in China, Shanghai Minimally Invasive Orthopedic Medical Technology Co., Ltd. The recall pertains to the Hip Joint Femoral Neck Component, specifically model PHAC1254, 8 Varus.The primary issue identified was an unexpectedly higher rate of femoral neck fractures in the affected product compared to other models. This critical defect could not be detected through visual inspection or other preoperative assessment methods, posing a potential risk to patients undergoing total hip arthroplasty. A total of 185 units were affected globally, with 164 units sold in China across all batches.Under the NMPA's regulatory oversight, the company initiated a Class II recall. The required corrective actions included immediately instructing distributors to cease sales of the identified product and actively recalling all affected units from the market based on product traceability information to prevent potential patient harm and ensure public safety.
- Company
- Microport Orthopedics, Inc.
ID · 0671d471-38aa-4985-983b-44e9c52ced41