China NMPA Product Recall - Electronic thoracoscopic and laparoscopic
Discuss this record with AI
Olympus Trading (Shanghai) Co., Ltd., acting for manufacturer Olympus Winter & The GmbH, initiated a voluntary Class II recall of its Electronic Thoracolaparoscopic Endoscopes, specifically models WA50040A and WA50042A. This action, reported to the National Medical Products Administration (NMPA) on November 16, 2015, and publicly announced on December 15, 2015, was prompted by complaints from overseas markets. These complaints detailed instances where the endoscope's tip experienced excessive temperatures during use. A subsequent investigation confirmed that a small number of temperature sensors in specific product batches manufactured after February 2015 might be faulty. This flaw could lead to unintended control of the heating element, causing the endoscope's tip to overheat. Such overheating presents a potential risk of thermal injury to tissues or organs if the device accidentally contacts them during surgical procedures. While no related adverse events were reported within China, the company undertook this global recall. In accordance with NMPA's recall management framework, affected medical institutions are mandated to immediately discontinue using the identified 184 units distributed in China and facilitate their return.