China NMPA Product Recall - Monitor/Defibrillator
Discuss this record with AI
Philips (China) Investment Co., Ltd. has initiated a voluntary Class II recall for its HeartStart MRx monitor/defibrillator devices. This recall, reported on March 11, 2020, via the National Medical Products Administration (NMPA) and the Shanghai Municipal Drug Administration, addresses a significant safety concern regarding device integrity.The core issue is that the HeartStart MRx may suffer internal damage from drops or severe mechanical impact, even without visible external damage or an immediate malfunction signal from the 'Ready-to-Use' indicator. A critical flaw is that the device will not immediately alert users to such internal damage, only flagging the problem during its next scheduled automatic self-test or operational check. This creates a safety risk unless the user performs an immediate manual check as per the instruction manual after any incident of dropping or mishandling.As a required action, Philips is recalling affected models, specifications, and batches of the monitor/defibrillator (Registration Certificate No.: 20163214004). The company urges users to strictly follow the instruction manual by manually inspecting the device immediately after any mechanical impact to ensure proper functionality and patient safety.
ID · f075eaca-cbf0-4965-bc68-20d668fecc0e