China NMPA Product Recall - Orthopedic surgical instruments - medullary reamer, perfusion, and suction system (trade name: RIA)
Discuss this record with AI
On December 22, 2014, Johnson & Johnson (Shanghai) Medical Devices Co., Ltd., in coordination with manufacturer Synthes GmbH, initiated a voluntary Class 1 recall of specific Orthopedic Surgical Instruments: the Reaming, Perfusion, and Suction System (Trade Name: RIA). This regulatory action was overseen by the National Medical Products Administration (NMPA) of China. The core issue identified was a discrepancy in the product's labeled expiration date. Stability and biocompatibility data only supported a two-year validity period from the manufacturing date, yet affected products were incorrectly labeled with a ten-year shelf-life. While no direct evidence confirmed adverse events, a theoretical risk of cytotoxicity after the two-year period was acknowledged due to insufficient data supporting the extended duration. The recall involved specific models, including 314.745S, 352.260S, and 357.718.02S, with 25 units imported into China. Crucially, all affected products were intercepted after import and had not been sold within the Chinese market, minimizing potential patient exposure. The required action was a comprehensive voluntary recall to rectify the mislabeling and ensure product validity aligns with scientific evidence.