China NMPA Product Recall - Maxillofacial surgical instruments (trade name: SYNTHES)
Discuss this record with AI
Johnson & Johnson (Shanghai) Medical Devices Co., Ltd., on behalf of manufacturer Synthes GmbH, initiated a voluntary Class II recall of specific maxillofacial surgical instruments, formally reported to the National Medical Products Administration (NMPA) on January 21, 2015. The recall addresses a product issue with the "beaked bending-cutting forceps" (Model 329.143, SYNTHES brand), which were found to fracture when used to bend or cut thicker mandibular bone plates. This deficiency, identified through complaint records, carries the risk of delaying surgical procedures should a replacement tool be needed during an operation. While the instrument is auxiliary and does not directly contact patients, the product's non-conformance posed an operational challenge, though no direct patient adverse events were reported. Under the NMPA's regulatory framework, a Class II recall indicates a remote probability of serious adverse health consequences. The required actions involve Johnson & Johnson notifying all affected distributors to immediately cease using the identified products and arrange for their return. All 41 units imported to China, encompassing those sold, consigned, and held as fixed assets, are to be either returned to the manufacturer or destroyed locally to prevent further use.
- Company
- Synthes GmbH
ID · 7fe0ed3f-a232-4f0a-80ef-df0a9e9f72ee