•Abaxis Inc•August 22, 2019

Abaxis Inc – FDA EIR, Aug 22, 2019

Record Details

An FDA follow-up inspection of Abaxis Inc., a medical device manufacturer in Union City, CA, was conducted from August 20-22, 2019. The inspection, carried out under Compliance Program 7382.845 for Medical Device Manufacturers, focused on the company's progress in addressing previous issues and compliance with regulatory standards for their Piccolo Xpress Point-Of-Care Blood Chemistry Analyzer and Rotors. The regulatory framework applied includes 21 CFR 820.198(d), 21 CFR 803.50(a)(1), and 21 CFR 820.20(c).

The inspection identified three main violations, resulting in an FDA Form 483. First, Abaxis failed to adequately identify and submit Medical Device Reports (MDRs) for events reasonably suggesting a marketed device may have caused or contributed to serious injury, specifically citing a complaint (18-0031H) where an erroneous high potassium result from the Piccolo device led to a patient's medical intervention and hospitalization. This MDR was not filed within the required 30 days. Second, management with executive responsibility had not reviewed the suitability and effectiveness of the quality system with sufficient frequency, with no review conducted since May 2018.

Required actions include management's promise to correct the identified deficiencies, including submitting the overdue MDR. Furthermore, the company reported ongoing efforts to file a new 510(k) by November 2019 for their Piccolo Potassium Assay, addressing a prior design change made without proper regulatory clearance. They had also previously revised procedures for Design Control, Document Change Control, and Risk Management in response to an April 2019 Warning Letter.

Company
Abaxis Inc
Inspection date
August 22, 2019
Product type
devices
Office
San Francisco District Office
People
  • James A. Lane (Investigator)
  • Eric J. Heinze (investigator)
View Company Profile

ID · c2460568-ad13-4dac-a9b3-87c7a8c52578

Violation Codes7
21 CFR 820.198(d)21 CFR 803.50(a)(1)21 CFR 820.20(c)21 CFR 80321 CFR 820.10021 CFR 820.3021 CFR 820.50

Full citation text and observation details available on the Dashboard.

Other documents for Abaxis Inc
  • 483 - 2019-08-22August 22, 2019
  • WARNING_LETTER - 2018-04-25April 25, 2018
  • 483 - 2008-08-08August 8, 2008

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Abaxis Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.