APN Health LLC – FDA EIR, Oct 10, 2019

Record Details

An FDA inspection of APN Health LLC, a Waukesha, WI-based medical device manufacturer, was conducted from October 1-10, 2019. The inspection focused on the company's Navik 3D cardiac mapping system, a Class II medical device, under the Quality System Inspection Technique (QSIT) and medical device regulations (21 CFR Part 820). The inspection resulted in two formal observations (Form FDA 483) and nine verbal observations, indicating significant quality system deficiencies.

The main written observations included inadequate design verification procedures and complaint handling processes. For design verification, APN Health LLC failed to consistently follow its "Test Case Creation and Execution" procedure for Navik 3D software version 2.2, with issues such as missing objective evidence, unaddressed failed test cases, and unfulfilled verification plan requirements. Regarding complaint handling, the firm initiated complaint records after software fixes were deployed, improperly combined multiple feedback forms, lacked documented Medical Device Report (MDR) evaluations, and prematurely closed complaints. Additionally, MDR evaluations did not adequately assess the potential for serious injury or death.

Verbal observations addressed further quality system concerns, including improving Corrective and Preventive Action (CAPA) scope, ensuring hazard alignment with complaints, documenting device history record reviews, revising computerized systems and supplier lifecycle procedures to include consultants and MDR/recall responsibilities, enhancing the MDR procedure, increasing internal audit frequency, and retaining all test data. Management committed to correcting all observations and providing a written response for the formal findings within 15 business days.

Company
APN Health LLC
Inspection date
October 10, 2019
Product type
devices
Office
Center for Devices and Radiological Health
People
  • David K. Geddam (recipient)
  • Jasbir Sra (recipient)
  • Michelle J. Glembin (investigator)
View Company Profile

ID · 4769a7c5-1b3f-4ff3-8774-a895d727170d

Violation Codes9
21 CFR 820.19821 CFR 820.3021 CFR 820.10021 CFR 820.3(k)21 CFR 807.9721 CFR 820.198(a)21 CFR 80321 CFR 820.30(f)21 CFR 806

Full citation text and observation details available on the Dashboard.

Other documents for APN Health LLC
  • 483 - 2023-09-01September 1, 2023
  • 483 - 2019-10-10October 10, 2019

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore APN Health LLC
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.