Record Details

An FDA Level 2 QSIT inspection was conducted at Pall Newquay Ltd. in Cornwall, United Kingdom, from January 30 to February 1, 2017. This routine inspection, part of the DMPTI FY-17 medical device workplan, covered all four major quality subsystems, focusing on the company's intravenous line filter, a Class II medical device. While no FDA-483 was issued, indicating no significant regulatory violations, the inspection identified several areas for improvement.

Discussions with management highlighted the need to revise the firm's training procedure (PNQP-6016A) to specify timeframes for new employee reviews and ensure accurate transcription on training records. Additionally, environmental monitoring procedures (SOP 10545, SOP 9760) needed updating to explicitly state that sampling occurs during manufacturing activities, aligning documentation with current practice. Suggestions were also made to enhance the organization and presentation of Corrective and Preventive Action (CAPA) records for better clarity.

Pall Newquay Ltd. management acknowledged these points and committed to implementing the necessary revisions, including updating SOPs and initiating non-conformance reports to address the identified discrepancies. The firm's recall procedure was also reviewed, noting a proactive development of a new procedure for non-reportable field corrections, aligning with regulatory expectations under 21 CFR 806.

Company
Pall Newquay Ltd.
Inspection date
February 1, 2017
Product type
devices
Office
Office of Regulatory Affairs
Person
  • Amy Cramer
View Company Profile

ID · e21ec3e3-ae83-4fe6-a884-ce1514b8ab20

Violation Codes3
21 CFR 820.3021 CFR 820.10021 CFR 806

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Pall Newquay Ltd.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.