FDA Inspection
Cardio Medical ProductsCardio Medical Products – FDA Inspection, Nov 5, 2019
Record Details
This FDA Inspection record concerns Cardio Medical Products, with an inspection on November 5, 2019, issued by the Center for Devices and Radiological Health, covering devices.
- Company
- Cardio Medical Products
- Inspection Date
- November 5, 2019
- Product Type
- Devices
View Company Profile
ID · db6706eb-fc98-4798-81a7-145a3fab54dc
Inspection Facts
Voluntary Action Indicated (VAI)
Objectionable conditions were found, but FDA is not prepared to take or recommend regulatory action.
- Inspection number
- 1108916
- FDA center
- Medical Devices & Rad Health
- Project area
- Compliance: Devices
FDA Citations
10- 21 CFR 820.40Procedures not adequately established or maintained
- 21 CFR 820.20(b)(3)Management representative
- 21 CFR 820.100(a)Lack of or inadequate procedures
- 21 CFR 820.30(a)Design control - no procedures
- 21 CFR 820.20(c)Management review - Lack of or inadequate procedures
- 21 CFR 820.198(a)Lack of or inadequate complaint procedures
- 21 CFR 820.25(b)Training - Lack of or inadequate procedures
- 21 CFR 820.50Purchasing controls, Lack of or inadequate procedures
- 21 CFR 820.30(j)Design history file
- 21 CFR 803.17Lack of Written MDR Procedures
Full citation text and observation details available on the Dashboard.
Other FDA inspections of Cardio Medical Products
1| Date | Classification | Office | Project area |
|---|---|---|---|
| February 1, 2024 | VAI | Center for Devices and Radiological Health | Compliance: Devices |
Other documents for Cardio Medical Products
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.