FDA Inspection
CAREstream Medical LtdCAREstream Medical Ltd – FDA Inspection, Feb 6, 2026
Record Details
This FDA Inspection record concerns CAREstream Medical Ltd, with an inspection on February 6, 2026, issued by the Florida District Office, covering medical devices & rad health.
- Company
- CAREstream Medical Ltd
- Inspection Date
- February 6, 2026
- Product Type
- Medical Devices & Rad Health
View Company Profile
ID · 2e9d2f50-5109-4578-829c-899034a0d502
Inspection Facts
Voluntary Action Indicated (VAI)
Objectionable conditions were found, but FDA is not prepared to take or recommend regulatory action.
- Inspection number
- 1301490
- FDA center
- Medical Devices & Rad Health
- Project area
- Compliance: Medical Devices
FDA Citations
7- 21 CFR 801.20(a)Label to bear a unique device identifier
- ISO 13485:2016 Clause 4.1.1QMS not documented and maintained adequately
- ISO 13485:2016 Clause 7.3.1Design and development procedures not documented
- ISO 13485:2016 Clause 7.4.1Records of supplier (re)evaluation, selection, and monitoring not maintained
- ISO 13485:2016 Clause 7.5.6Computer software used in production and service provision not validated
- ISO 13485:2016 Clause 8.2.2Complaint handling procedures not documented
- ISO 13485:2016 Clause 5.2Top management did not ensure requirements are determined and met
Full citation text and observation details available on the Dashboard.
Other FDA inspections of CAREstream Medical Ltd
1| Date | Classification | Office | Project area |
|---|---|---|---|
| April 27, 2022 | NAI | Center for Biologics Evaluation and Research | Human Cellular, Tissue, and Gene Therapies |
Other documents for CAREstream Medical Ltd
- FDA Inspection 1168008 - 2022-04-27
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.