FDA Inspection
Genesis BPS, LLC.Genesis BPS, LLC. – FDA Inspection, Nov 20, 2017
Record Details
This FDA Inspection record concerns Genesis BPS, LLC., with an inspection on November 20, 2017, issued by the Center for Devices and Radiological Health, covering devices.
- Person
- Melissa A. Zuppe
View Company Profile
ID · 5ba1d989-5a15-4f14-a67f-26f59251339a
Inspection Facts
Voluntary Action Indicated (VAI)
Objectionable conditions were found, but FDA is not prepared to take or recommend regulatory action.
- Inspection number
- 1031705
- FDA center
- Medical Devices & Rad Health
- Project area
- Compliance: Devices
- Outcome
- Form 483 issued
FDA Citations
4- 21 CFR 820.198(a)Complaints
- 21 CFR 820.40(a)Document review, approval by designated individual
- 21 CFR 820.30(g)Design validation - production units
- 21 CFR 803.17Lack of Written MDR Procedures
Full citation text and observation details available on the Dashboard.
Other FDA inspections of Genesis BPS, LLC.
4| Date | Classification | Office | Project area |
|---|---|---|---|
| July 31, 2025 | NAI | Center for Devices and Radiological Health | Postmarket Assurance: Devices |
| October 5, 2021 | NAI | Center for Devices and Radiological Health | Compliance: Devices |
| June 10, 2015 | VAI | Center for Devices and Radiological Health | Compliance: Devices |
| February 11, 2013 | OAI | Center for Devices and Radiological Health | Compliance: Devices |
Other documents for Genesis BPS, LLC.
- FDA Inspection 1281312 - 2025-07-31
- FDA Inspection 1154238 - 2021-10-05
- FDA Inspection 930793 - 2015-06-10
- FDA Inspection 819283 - 2013-02-11