Official Action Indicated (OAI)
Significant objectionable conditions were found and regulatory or administrative action is recommended.
- Inspection number
- 1058645
- FDA center
- Medical Devices & Rad Health
- Project area
- Compliance: Devices
- Outcome
- Form 483 issued
FDA Citations
7- 21 CFR 820.100(a)Lack of or inadequate procedures
- 21 CFR 820.30(a)Design control - no procedures
- 21 CFR 820.198(a)Lack of or inadequate complaint procedures
- 21 CFR 820.25(b)Training - Lack of or inadequate procedures
- 21 CFR 820.50Purchasing controls, Lack of or inadequate procedures
- 21 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
- 21 CFR 803.17Lack of Written MDR Procedures
Full citation text and observation details available on the Dashboard.
Other FDA inspections of Heartware, Inc.
4| Date | Classification | Office | Project area |
|---|---|---|---|
| May 4, 2016 | OAI | Center for Devices and Radiological Health | Compliance: Devices |
| January 24, 2014 | OAI | Center for Devices and Radiological Health | Compliance: Devices |
| August 30, 2013 | NAI | Center for Devices and Radiological Health | Compliance: Devices |
| July 20, 2011 | VAI | Center for Devices and Radiological Health | Compliance: Devices |
Other documents for Heartware, Inc.
- FDA Inspection 971984 - 2016-05-04
- FDA Inspection 864927 - 2014-01-24
- FDA Inspection 851937 - 2013-08-30
- FDA Inspection 738736 - 2011-07-20