FDA Inspection
Northern Digital Inc.Northern Digital Inc. – FDA Inspection, Mar 28, 2013
Record Details
This FDA Inspection record concerns Northern Digital Inc., with an inspection on March 28, 2013, issued by the Center for Devices and Radiological Health, covering devices.
- Company
- Northern Digital Inc.
- Inspection Date
- March 28, 2013
- Product Type
- Devices
View Company Profile
ID · b9347f5f-28b4-4e88-8774-0b0dca6bced2
Inspection Facts
Official Action Indicated (OAI)
Significant objectionable conditions were found and regulatory or administrative action is recommended.
- Inspection number
- 824249
- FDA center
- Medical Devices & Rad Health
- Project area
- Compliance: Devices
FDA Citations
7- 21 CFR 820.30(f)Design verification - output does not meet input requirement
- 21 CFR 820.20(a)Management ensuring quality policy is understood
- 21 CFR 820.100(a)Lack of or inadequate procedures
- 21 CFR 820.198(a)Lack of or inadequate complaint procedures
- 21 CFR 820.75(a)Documentation
- 21 CFR 820.180Availability
- 21 CFR 803.17(b)(1)Info evaluated to determine if event was reportable
Full citation text and observation details available on the Dashboard.
Other FDA inspections of Northern Digital Inc.
3| Date | Classification | Office | Project area |
|---|---|---|---|
| August 28, 2025 | OAI | Office of Inspections and Investigations | Compliance: Medical Devices |
| August 14, 2024 | NAI | Center for Devices and Radiological Health | Postmarket Assurance: Devices |
| July 9, 2015 | NAI | Center for Devices and Radiological Health | Postmarket Assurance: Devices |
Other documents for Northern Digital Inc.
- FDA Inspection 1279866 - 2025-08-28
- FDA Inspection 1249033 - 2024-08-14
- FDA Inspection 933943 - 2015-07-09
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.