VAX-D Medical Technologies LLCVAX-D Medical Technologies LLC – FDA Inspection, Apr 13, 2012
Record Details
This FDA Inspection record concerns VAX-D Medical Technologies LLC, with an inspection on April 13, 2012, issued by the Center for Devices and Radiological Health, covering devices.
- Inspection Date
- April 13, 2012
- Product Type
- Devices
View Company Profile
ID · aadb7654-1981-403a-9330-b2ec46803ad0
Inspection Facts
Voluntary Action Indicated (VAI)
Objectionable conditions were found, but FDA is not prepared to take or recommend regulatory action.
- Inspection number
- 776703
- FDA center
- Medical Devices & Rad Health
- Project area
- Postmarket Assurance: Devices
FDA Citations
12- 21 CFR 820.181DMR - not or inadequately maintained
- 21 CFR 820.40Lack of procedures, or not maintained
- 21 CFR 820.100(a)Lack of or inadequate procedures
- 21 CFR 820.20(c)Management review - Lack of or inadequate procedures
- 21 CFR 820.198(a)Lack of or inadequate complaint procedures
- 21 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
- 21 CFR 820.50Purchasing controls, Lack of or inadequate procedures
- 21 CFR 820.198(e)Records of complaint investigation
- 21 CFR 820.184DHR content
- 21 CFR 806.20(b)(2)Device model identification
Showing 10 of 12 citations.
Full citation text and observation details available on the Dashboard.
Other FDA inspections of VAX-D Medical Technologies LLC
1| Date | Classification | Office | Project area |
|---|---|---|---|
| February 5, 2013 | VAI | Center for Devices and Radiological Health | Compliance: Devices |
Other documents for VAX-D Medical Technologies LLC
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.