Record Details

The FDA issued a Warning Letter to Edward O’Neill after a May 2022 inspection of their Ottawa, Canada facility, regarding the i-STAT cTnI Test. The device, for cardiac troponin I measurement, was found adulterated under section 501(f)(1)(B) of the Act due to lacking an approved Premarket Approval (PMA) or Investigational Device Exemption (IDE). It was also misbranded under section 502(o) for failing to submit a 510(k) for significant design modifications, as required by section 510(k) and 21 CFR 807.81(a)(3).

Since 2003, the firm made multiple significant design changes to the i-STAT cTnI cartridge, including material, reagent, and sensor chip modifications. These changes significantly altered risks and could affect device safety/effectiveness, requiring new 510(k) submissions, which were not provided, rendering the current cartridge unapproved.

Devices were also adulterated under section 501(h) for non-conformity with Quality System (QS) regulation (21 CFR Part 820). Violations included inadequate procedures for design change identification, documentation, validation, and approval (21 CFR 820.30(i)), as cumulative effects were not assessed. Additionally, the firm failed to adequately maintain accessible records (21 CFR 820.180), exemplified by missing original design files. The firm's FDA 483 response was deemed inadequate. Prompt corrective actions, a 15-business-day written response, and a systemic review of all device changes are mandated to prevent further regulatory action and business implications.

Company
Abbott Point of Care Canada Limited
Inspection date
May 19, 2022
Product type
Devices
Office
Center for Devices and Radiological Health
People
  • Kerry Caldwell
  • Timothy T. Stenzel (Director)
View Company Profile

ID · 7b8c8900-28e8-43f0-835e-9aada047b889

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(f)(1)(B)21 U.S.C. 360e(a)21 U.S.C. 360j(g)21 U.S.C. 352(o)21 U.S.C. 360(k)21 CFR 807.81(a)(3)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(i)

Full citation text and observation details available on the Dashboard.

Other documents for Abbott Point of Care Canada Limited
  • 483 - 2022-05-16May 16, 2022

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Abbott Point of Care Canada Limited
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.